The tissue-equivalence of the Alderson Rando anthropomorphic phantom for X-rays of diagnostic qualities
- 1 January 1981
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 26 (1) , 133-139
- https://doi.org/10.1088/0031-9155/26/1/013
Abstract
Comparative measurements were made of depth-dose profiles in an Alderson Rando phantom and a similarly shaped phantom of water which closely simulates the attenuation properties of human muscle. Measurements were carried out at three different kV settings of the X-ray generator, covering the lower-energy range of normal diagnostic practice over which attenuation differences are likely to be most important. The results show that whilst the Alderson phantom can undoubtedly be extremely useful in medical dosimetry, corrections to the dose distributions observed for diagnostic qualities of radiation may sometimes be necessary to allow for departures from tissue-equivalence.Keywords
This publication has 3 references indexed in Scilit:
- Tissue substitutes in experimental radiation physicsMedical Physics, 1978
- A TLD system based on lithium borate for the measurement of doses to patients undergoing medical irradiationPhysics in Medicine & Biology, 1976