EFFECT OF SMALL AMOUNTS OF I-131 ON TC-99M DETECTION BY SCINTILLATION CAMERA
- 1 January 1976
- journal article
- research article
- Vol. 17 (12) , 1100-1101
Abstract
A patient who recently received a therapeutic dose of Na131I showed no 99mTc activity on the scintillation camera when a pertechnetate brain scan was attempted. Subsequent experiments showed that, when even small amounts of 131I were present in the field of view of a scintillation camera equipped with a collimator designed for low-energy photons, counting rates from 99mTc were markedly affected.This publication has 0 references indexed in Scilit: