Theoretical Formulation of Sieving Coefficient Evaluation for Membrane Plasma Separation
- 1 August 1985
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 9 (3) , 250-254
- https://doi.org/10.1111/j.1525-1594.1985.tb04387.x
Abstract
Three different sieving coefficient definitions were theoretically investigated. It has been shown that the local sieving coefficient, characterizing membrane property, can be calculated under specified assumptions from input — output measurements. In contrast, the evaluation of the device sieving coefficient, useful in kinetic modeling, is validated in a situation in which calculation of the local sieving coefficient is connected with substantial error. The relationship between different sieving coefficient definitions is also given.Keywords
This publication has 3 references indexed in Scilit:
- Characterization of Polyethylene Membrane for Plasma SeparationArtificial Organs, 1983
- SOLUTE FLUX IN HEMODIALYSIS AND HEMOFILTRATION MEMBRANESAsaio Journal, 1977
- The theory of measurement of osmotic pressureRecueil des Travaux Chimiques des Pays-Bas, 1951