Evaluation of a new homogeneous enzyme inhibitor immunoassay of serum thyroxine with use of a bichromatic analyzer.
- 1 November 1980
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 26 (12) , 1723-1726
- https://doi.org/10.1093/clinchem/26.12.1723
Abstract
We evaluated a new homogeneous enzyme immunoassay of thyroxine with use of a discrete analyzer (the ABA-100 Bichromatic Analyzer), modified with an auxiliary dispenser assembly. The assay is based on inhibition of hydrolysis of the substrate, acetyl-beta-(methylthio)choline iodide, by acetylcholinesterase (acetylcholine hydrolase; EC 3.1.1.7). Thyroxine covalently linked to a cholinesterase inhibitor irreversibly inhibits acetylcholinesterase, but if this thyroxine conjugate is bound to antibody it is not inhibitory. Seventy-five patients' samples may be analyzed in 1 h of instrument time. Precision and accuracy are excellent, results correlate well with those by radioimmunoassay, and there were no instances of confused clinical interpretation resulting from use of the proposed assay.This publication has 5 references indexed in Scilit:
- Thyroid function screening in psychiatric patientsJAMA, 1979
- The diagnostic dilemma of isolated hyperthyroxinemia in acute illnessJAMA, 1979
- Misleadingly Low Free Thyroxine Index and Usefulness of Reverse Triiodothyronine Measurement in Nonthyroidal IllnessesAnnals of Internal Medicine, 1979
- The Free Triiodothyronine (T_3) IndexAnnals of Internal Medicine, 1978
- Evaluation of homogeneous enzyme immunoassay of serum thyroxine with use of bichromatic analysis.Clinical Chemistry, 1978