Gas-chromatographic determination of 5-fluorocytosine in human serum.
Open Access
- 1 June 1976
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 22 (6) , 772-776
- https://doi.org/10.1093/clinchem/22.6.772
Abstract
We describe a sensitive and precise gas-chromatographic method, in which cytosine is used as the internal standard, for determination of an antifungal agent, 5-fluorocytosine, in serum. The trimethylsilyl derivative of this drug is well separated from the internal standard and from normal serum constituents. Amphotericin B does not interfere with the determination of 5-fluorocytosine. The lower limit of detection for 5-fluorocytosine is 1 mg/liter when 200 mul of serum is analyzed. Within-run precision (CV), established by analysis of 10 replicates, was 4.5% at a concentration of 19.9 mg/liter. Twenty-five serum samples were analyzed for 5-fluorocytosine by a microbiological assay and by the gas-chromatographic method. Mean value observed with the bioassay was 78.5 mg/liter and with our procedure was 69.4 mg/liter. When values for our assay were regressed against values for the bioassay, slope of the least-squares line was 0.85, intercept was 2.7 mg/liter, and r was 0.93.This publication has 3 references indexed in Scilit:
- 5-Fluorocytosine in the Treatment of Mycotic InfectionsAnnals of Internal Medicine, 1972
- 5-Fluorocytosine: An Oral Antifungal CompoundAnnals of Internal Medicine, 1972
- 5-Fluorocytosine in the Treatment of Cryptococcal and Candida MycosesAnnals of Internal Medicine, 1971