Pitfalls in respiratory monitoring of premature infants during kangaroo care.
Open Access
- 1 March 1995
- journal article
- research article
- Published by BMJ in Archives of Disease in Childhood: Fetal & Neonatal
- Vol. 72 (2) , F115-F117
- https://doi.org/10.1136/fn.72.2.f115
Abstract
The reliability of respiratory monitoring, using either chest or back electrodes, was studied in 13 preterm infants during kangaroo care (infant-parent skin to skin contact). In three out of four infants with chest electrodes both infant and parental respiration were clearly visible on pneumograms. In these infants apnoeic pauses were not registered because parental respiration was recorded as infant breathing. Bradycardia and oxygen saturation were, however, properly registered. In infants with electrodes placed on the back infant respiration was less superimposed by parental breathing. However, even in some of these infants parental respiration was visible in the pneumograms. It is concluded that during kangaroo care the electrodes should be placed on the back and monitoring should always include heart rate and oxygen saturation.Keywords
This publication has 7 references indexed in Scilit:
- Reconciliation and healing for mothers through skin-to-skin contact provided in an American tertiary level intensive care nursery.1993
- Current knowledge about skin-to-skin (kangaroo) care for preterm infants.1991
- Cardiorespiratory monitoring.1991
- Cardiorespiratory MonitoringClinics in Perinatology, 1991
- Physiological Effects of Kangaroo Care in Very Small Preterm InfantsNeonatology, 1991
- Skin to skin contact for very low birthweight infants and their mothers.Archives of Disease in Childhood, 1988
- Apnea Monitor Failure in Infants With Upper Airway ObstructionPediatrics, 1977