An Effective Algorithm for Extracting Serial Endocardial Borders from 2-Dimensional Echocardiograms
- 1 June 1984
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. BME-31 (6) , 441-447
- https://doi.org/10.1109/tbme.1984.325359
Abstract
An effective image processing algorithm for extracting the endocardial border from a series of 2-dimensional echocardiograms (2D echo) is proposed. The main steps involved in this algorithm are coordinate transformation, regional co-occurrence matrix thresholding, search region limiting, border detecting, and spatial and temporal smoothing. Using these steps, some major difficulties related to 2D echo image processing including low signal-to-noise ratio, echo drop out, and gray level intensity variability have been dramatically reduced. All endocardial borders in an entire cardiac cycle are computed. This is done automatically on the basis of observer-drawn outlines at end systole and end diastole. These outlines facilitate the quantitative analysis by defining a range in which both boundary dimension and shape changes should occur. Image processing is restricted to this range. The validation studies have been perfonned on test balloons where both image quality has been changed and complex motion abnormalities have been experimentally produced. Good results have also been obtained in patient studies.Keywords
This publication has 8 references indexed in Scilit:
- Automatic boundary detection of the left ventricle from cineangiogramsPublished by Elsevier ,2003
- Reproducibility of left ventricular volumes by two-dimensional echocardiographyJournal of the American College of Cardiology, 1983
- An Approach to Optimal Threshold Selection on a Sequence of Two-Dimensional Echocardiographic ImagesIEEE Transactions on Biomedical Engineering, 1982
- Computerized image processing in cross-sectional echocardiographyThe American Journal of Cardiology, 1980
- Assessment of left ventricular ejection fraction and volumes by real-time, two-dimensional echocardiography. A comparison of cineangiographic and radionuclide techniques.Circulation, 1979
- Three-dimensional echocardiograms and two-dimensional echocardiographic images at desired planes by a computerized systemUltrasound in Medicine & Biology, 1977
- Straight-Line Approximation for the Boundary of the Left Ventricular Chamber from a Cardiac CineangiogramIEEE Transactions on Biomedical Engineering, 1973
- On-line processing of the video-image for left ventricular volume determinationComputers and Biomedical Research, 1971