Gold-Coated NIR Stents in Porcine Coronary Arteries
Open Access
- 23 January 2001
- journal article
- other
- Published by Wolters Kluwer Health in Circulation
- Vol. 103 (3) , 429-434
- https://doi.org/10.1161/01.cir.103.3.429
Abstract
Background —As endovascular stents are altered to add functionality, eg, by adding radiopaque coatings, biocompatibility may suffer. Methods and Results —We examined the vascular response in porcine coronary arteries to stainless steel gold-coated NIR stents (7-cell, Medinol, Inc). Stents, 9 and 16 mm in length, were left bare or coated with a 7-μm layer of gold. Physical and material effects were examined in four different gold-coated stent types, two at each length that either had the coating applied to the standard strut, ie, gold coated thicker than controls, or had the coating applied to thinned struts, ie, gold coated of the same thickness as control struts. Simple gold coating exacerbated intimal hyperplastic and inflammatory reactions over 28 days, but postplating thermal processing smoothed the coating surface and negated the adverse tissue response to gold. The relative amounts of base steel and gold coating and their resistances to expansion and collapse determined the extent of stent recoil. Conclusions —Gold coatings enhance the radiopacity of steel stents, but not without effects on vascular repair. Material effects predominate and can be abrogated by heating coated stents to alter surface finish and material purity. Clinical results may suffer unless consideration is given to material and physical effects of gold.Keywords
This publication has 11 references indexed in Scilit:
- Rapid in vitro biocompatibility assay of endovascular stents by flow cytometry using platelet activation and platelet-leukocyte aggregationCytometry, 1999
- The dawn of dentistry: dentistry among the EtruscansInternational Dental Journal, 1997
- The influence of dental metal alloys on cell proliferation and fibronectin arrangement in human fibroblast culturesArchives of Oral Biology, 1997
- Metal implants and surface reactionsInjury, 1996
- Early Eyelid Rehabilitation in Facial Nerve ParalysisThe Laryngoscope, 1996
- Comparison between First-Generation (Fixed-Caliber) and Second-Generation (Self-Expanding, Large Caliber) Temporary Prostatic StentsUrologia Internationalis, 1996
- Hydrophilic surface modification of metallic endoluminal stentsJournal of Vascular Surgery, 1995
- Reaction of the aortic wall to six metallic stent materialsAcademic Radiology, 1995
- Encrustation of Intraprostatic Stents–a Comparative StudyBritish Journal of Urology, 1992
- Restenosis and the proportional neointimal response to coronary artery injury: Results in a porcine modelJournal of the American College of Cardiology, 1992