Mechanism of Obstruction of Closed-wound Suction Tubing
- 1 May 1979
- journal article
- research article
- Published by American Medical Association (AMA) in Archives of Surgery
- Vol. 114 (5) , 614-615
- https://doi.org/10.1001/archsurg.1979.01370290064013
Abstract
• The mechanism of obstruction of closed-wound drainage tubing was investigated by means of coagulation tests performed on wound drainage fluid and by examination of the contents of the tubes after their removal. Although clotting is commonly thought to be responsible for the obstruction, the wound drainage fluid was found to be essentially incoagulable and little fibrin was evident within the tubes. By contrast, bits of tissue were frequently found within the tubes, and these frequently virtually occluded the lumen. This observation, that tissue fragments are responsible for tube obstruction, permits a rational approach to the solution of this problem. For example, meticulous wound flushing and irrigation, or perhaps tubing of different design, might lead to a reduced incidence of tube failure. (Arch Surg 114:614-615, 1979)This publication has 3 references indexed in Scilit:
- Comparison of the Reaction-rate and Clot-density Methods for Determination of Plasma FibrinogenAmerican Journal of Clinical Pathology, 1977
- RAPID LATEX-SCREENING TEST FOR DETECTION OF FIBRIN/FIBRINOGEN DEGRADATION PRODUCTS IN URINE AFTER RENAL TRANSPLANTATIONThe Lancet, 1973