A comparison of the forces exerted during laryngoscopy
- 1 October 1996
- journal article
- clinical trial
- Published by Wiley in Anaesthesia
- Vol. 51 (10) , 912-915
- https://doi.org/10.1111/j.1365-2044.1996.tb14955.x
Abstract
The forces exerted at laryngoscopy with the McCoy and Macintosh blades have been compared in 40 patients. The variables measured were the duration of laryngoscopy, the three maximally-applied forces and the mean force. The mean (SD) forces recorded were 18.9 (7.82) and 10.1 (5.33) N, respectively, with the Macintosh and the McCoy blades (p < 0.001) with a similar duration of laryngoscopy. There was a significant positive correlation between the mean force and patients' weight, height and body mass index for the Macintosh blade but only for weight and height for the McCoy blade. It is concluded that the use of the McCoy blade results in significantly less force being applied during laryngoscopy. This may be the reason for the reduction in the stress response reported previously with the use of the McCoy blade.Keywords
This publication has 21 references indexed in Scilit:
- Another trachea intubated with the McCoy laryngoscopeAnaesthesia, 1995
- The McCoy levering laryngoscope bladeAnaesthesia, 1994
- The McCoy levering laryngoscope bladeAnaesthesia, 1994
- The McCoy levering laryngoscope bladeAnaesthesia, 1994
- The levering laryngoscopeAnaesthesia, 1993
- CARDIOVASCULAR AND CATECHOLAMINE RESPONSES TO LARYNGOSCOPY WITH AND WITHOUT TRACHEAL INTUBATIONBritish Journal of Anaesthesia, 1987
- Laryngoscope Handle MalfunctionAnesthesiology, 1986
- PLASMA CATECHOLAMINE RESPONSES TO TRACHEAL INTUBATIONBritish Journal of Anaesthesia, 1983
- Circulatory Changes during Direct Laryngoscopy and Tracheal IntubationAnesthesiology, 1977
- ENDOTRACHEAL INTUBATIONAnesthesiology, 1960