Exit Doses for Lung Fields measured by Lithium Fluoride Thermoluminescence
- 1 October 1965
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 38 (454) , 785-790
- https://doi.org/10.1259/0007-1285-38-454-785
Abstract
Lithium fluoride thermoluminescence dosimetry powder has been used in gelatine capsules of 10 mm length and 4 mm diameter to measure the exit doses in patients having treatment to the thorax by means of 8 MV X-rays. It was found that exit doses for fields passing through the lungs were up to 60% higher than the doses calculated on the assumption of homogeneous unit density tissue; the dose received at the tumor would thus be increased by about half this value. Such increases can be satisfactorily explained because the beam has passed through lung tissue containing air instead of unit density tissue. Where growth or fluid or fibrosis was present in the lung, the excess dose was less marked, and the exit dose became nearer to that expected for unit density tissue. In most cases it was possible to explain the observed exit doses, but the magnitude of the variation could not have been assessed from radiographs and clinical examination. It is therefore not possible to apply general correction factors, without exit dose measurement (or quantitative tomography) on each patient (Holodny, Gian, Ragazani, Bronstein and Laughlin, 1964; Fedoruk and Johns, 1957). Lithium fluoride powder makes it possible to make a large number of measurements on patients during routine treatment. Such measurements may prove useful for more precise estimates of the dose given in radiotherapeutic procedures, especially where results of lung treatments are being analysed with respect to dose.This publication has 19 references indexed in Scilit:
- Entrance and Exit Absorbed Dose Characteristics for a 6-Mev X-Ray GeneratorRadiology, 1964
- Thermoluminescent Radiation Dosimetry Utilizing LiFHealth Physics, 1964
- Radiation Dosimeter Utilizing the Thermoluminescence of Lithium FluorideScience, 1961
- II. The Effects of Some Variables on Arc Therapy Dosage DistributionsThe British Journal of Radiology, 1958
- Exit Dosage For 2-MVP X-RaysRadiology, 1958
- Measured Dose Distributions in Arc and Rotation Therapy: A Critical Comparison of Moving and Fixed Field TechniquesThe British Journal of Radiology, 1957
- Transmission Dose Measurement for Cobalt 60 Radiation with Special Reference to Rotation TherapyThe British Journal of Radiology, 1957
- II. The Patient and the Röntgen. Part IThe British Journal of Radiology, 1955
- Routine measurement of the rectal dose in the treatment of carcinoma of the cervix by radiumJournal of the Faculty of Radiologists, 1954
- Surface and Emergent Doses in RadiotherapyThe British Journal of Radiology, 1948