Quantitation in Positron Emission Computed Tomography
- 1 October 1982
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Computer Assisted Tomography
- Vol. 6 (5) , 987-999
- https://doi.org/10.1097/00004728-198210000-00022
Abstract
The intrinsic resolution (FWHM) of coincident bismuth germanate detector pairs was measured for geometries consistent with positron emission computed tomography systems designed for brain imaging. The resolutions were measured as a function of the position in the field of view (FOV) for single slice and multislice configurations for both the image and axial resolutions. System diameters of 66-42 cm were simulated, and intrinsic resolution variation for image and axial dimensions for a 24 cm FOV were 14 and 32% and 24 and 39%, respectively. Measurement of axial resolutions of interplane slices of multiplane systems employing interplane septa showed resolution variations for the 24 cm FOV of as large as a factor of 3. A simple model predicted the dominant features of the line shapes. Images of phantoms with small axial dimensions indicate that the variations in resolution, particularly in interplane data, seriously compromise the validity of point to point comparison of information within an image or between single plane and interplane images. Serious variations in axial resolution created artifacts and degraded the image resolution.This publication has 7 references indexed in Scilit:
- Quantitation in Positron Emission Computed Tomography: 5. Physical–Anatomical EffectsJournal of Computer Assisted Tomography, 1981
- In vivo quantitation of regional myocardial blood flow by positron-emission computed tomography.Circulation, 1981
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1980
- METHOD FOR OPTIMIZING SIDE SHIELDING IN POSITRON-EMISSION TOMOGRAPHS AND FOR COMPARING DETECTOR MATERIALS1980
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1979
- BISMUTH GERMANATE AS A POTENTIAL SCINTILLATION DETECTOR IN POSITRON CAMERAS1977
- SLICE GEOMETRY IN COMPUTER-ASSISTED TOMOGRAPHY1977