Sensitivity Calibration of Multichannel Raman Spectrometers Using the Least-Squares-Fitted Fluorescence Spectrum of Quinine
- 1 January 1988
- journal article
- other
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 42 (1) , 12-14
- https://doi.org/10.1366/0003702884428608
Abstract
A method of sensitivity calibration of multichannel Raman spectrometers is described which employs the fluorescence spectrum of quinine as an intensity standard. The reported absolute fluorescence spectrum of quinine was least-squares fitted with a polynomial so that it gave an interpolation of the spectrum in the “Raman scale.” The obtained least-squares-fitted fluorescence spectrum was cross-checked with the rotational Raman spectrum of D2. The fluorescence intensities calculated for every 100 cm−1 from the two Ar+ laser lines are tabulated for convenience.Keywords
This publication has 4 references indexed in Scilit:
- Relative Intensity Calibration of Rotational Raman SpectraApplied Spectroscopy, 1976
- Modified Technique for Determining the Wavelength-Sensitivity Curve of a SpectrofluorimeterApplied Optics, 1975
- SENSITIVITY CALIBRATION OF A RAMAN SPECTROMETER USING THE ROTATIONAL RAMAN SPECTRA OF DEUTERIUMChemistry Letters, 1974