Technique Factors and Image Quality as Functions of Patient Weight at Abdominal CT
Top Cited Papers
- 1 November 2000
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 217 (2) , 430-435
- https://doi.org/10.1148/radiology.217.2.r00nv35430
Abstract
To investigate how changes in kilovolt peak and milliampere second settings, and patient weight affect transmitted x-ray energy fluence and the image contrast-to-noise ratio (CNR) at abdominal computed tomography (CT). Cylinders of water were used as patient models, and x-ray spectra, including x-ray tube potentials of 80-140 kVp, were investigated. The mean photon energy and energy fluence transmitted through water cylinders with varying diameters and the image contrast for fat, muscle, bone, and iodine relative to water were determined. The effect of changing the x-ray tube potential on CNR also was investigated. At a constant kVp, increasing patient weight from 10 kg to 120 kg reduced the transmitted energy fluence by two orders of magnitude. Changing the x-ray tube potential from 80 kVp to 140 kVp increased the mean photon energy from approximately 52 keV to approximately 72 keV and thus reduced the image contrast relative to water by 12% for muscle, 21% for fat, 39% for bone, and 50% for iodine (approximate reduction values). Increasing the x-ray tube potential from 80 kVp to 140 kVp increased the CNR by a factor of 2.6 for muscle and by a factor of 1. 4 for iodine. With changes in patient weight at abdominal CT, x-ray tube potentials must be varied to maintain a constant detector energy fluence. Increasing the x-ray tube potential generally improves CNR.Keywords
This publication has 19 references indexed in Scilit:
- Effective doses to patients undergoing thoracic computed tomography examinationsMedical Physics, 2000
- Radiation Effective Doses to Patients Undergoing Abdominal CT ExaminationsRadiology, 1999
- CT scanning of the paranasal sinuses--the effect of reducing mAs.The British Journal of Radiology, 1997
- An approach for the estimation of effective radiation dose at CT in pediatric patients.Radiology, 1997
- Mottle on computed radiographs of the chest in pediatric patients.Radiology, 1996
- CT of the chest: minimal tube current required for good image quality with the least radiation dose.American Journal of Roentgenology, 1995
- Radiation dose reduction in CT of the pediatric pelvis.Radiology, 1994
- The effect of mAs variation upon computed tomography image quality as evaluated by in vivo and in vitro studies.Radiology, 1981
- Principles of computer assisted tomography (CAT) in radiographic and radioisotopic imagingPhysics in Medicine & Biology, 1976
- Photon cross sections, attenuation coefficients, and energy absorption coefficients from 10 keV to 100 GeVPublished by National Institute of Standards and Technology (NIST) ,1969