A robust inexpensive torque magnetometer with on-line data analysis
- 1 January 1982
- journal article
- Published by IOP Publishing in Journal of Physics E: Scientific Instruments
- Vol. 15 (1) , 128-131
- https://doi.org/10.1088/0022-3735/15/1/025
Abstract
A simple and robust torque magnetometer which has a wide sensitivity range, from 10-6 to 0.2 N m, is described. The design is based on that of Aldenkamp et al. (1960) but uses a modified mutual inductance method for measurement of angular displacement which in conjunction with a lock-in amplifier provides a direct reading of torque. This output is fed, together with a voltage proportional to magnet angle, to an on-line microcomputer, Fourier analysis of the torque curve, including a shearing correction takes approximately one minute following the complete input of data. The instrument is suitable for work down to liquid helium temperatures and up to about 450K.Keywords
This publication has 3 references indexed in Scilit:
- A double axis X-ray diffractometer for magnetostriction measurementsJournal of Physics E: Scientific Instruments, 1980
- The Magnetocrystalline Anisotropy of GadoliniumProceedings of the Physical Society, 1962
- Frictionless Recording Torque MagnetometerReview of Scientific Instruments, 1960