Performance evaluation of whole‐body NMR scanner antenna systems
- 1 August 1985
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 2 (4) , 328-335
- https://doi.org/10.1002/mrm.1910020403
Abstract
Whole‐body NMR imaging antennas (probes) are strongly affected by the inevitable magnetic and controllable dielectric losses. Using a convenient parallelepiped model, the magnetic losses are evaluated. By introducing the “ideal power gain,” the best possible antenna performance is delimited as a function of the frequency and the patient examined. By using a cylindrical model and introducing the “electrostatic quality factor” (EQF), the dielectric loss of any antenna can be estimated. With the help of the wideband field measurer described, a simple experimental arrangement, and the analysis developed, the performance of whole‐body NMR antennas can be compared and/or evaluated. An example is given for the application of this model at 6.4, 21, and 64 MHz for three typical patients.Keywords
This publication has 6 references indexed in Scilit:
- Preliminary experimental results in humans and animals with a superconducting, whole-body, nuclear magnetic resonance scanner.Radiology, 1982
- Nuclear magnetic resonance whole-body imager operating at 3.5 KGauss.Radiology, 1982
- An efficient decoupler coil design which reduces heating in conductive samples in superconducting spectrometersJournal of Magnetic Resonance (1969), 1979
- Radiofrequency losses in NMR experiments on electrically conducting samplesJournal of Magnetic Resonance (1969), 1979
- The sensitivity of the zeugmatographic experiment involving human samplesJournal of Magnetic Resonance (1969), 1979
- RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imagingPhysics in Medicine & Biology, 1978