Computer-assisted detection of pulmonary nodules: evaluation of diagnostic performance using an expert knowledge-based detection system with variable reconstruction slice thickness settings
- 2 December 2004
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 15 (2) , 203-212
- https://doi.org/10.1007/s00330-004-2544-5
Abstract
The purpose of this study was to evaluate the performance of a computer-assisted diagnostic (CAD) tool using various reconstruction slice thicknesses (RST). Image data of 20 patients undergoing multislice CT for pulmonary metastasis were reconstructed at 4.0, 2.0 and 0.75 mm RST and assessed by two blinded radiologists (R1 and R2) and CAD. Data were compared against an independent reference standard. Nodule subgroups (diameter >10, 4–10, U test. CAD was outperformed by readers at 4.0 mm (Az = 0.18, 0.62 and 0.69 for CAD, R1 and R2, respectively; PPP<0.05). The CAD advantage was best for nodules <10 mm (detection rates = 93.3, 89.9, 47.9 and 47.9% for R1 + CAD, R2 + CAD, R1 and R2, respectively). CAD using 0.75 mm RST outperformed radiologists in nodules below 10 mm in diameter and should be used to replace a second radiologist. CAD is not recommended for 4.0 mm RST.Keywords
This publication has 38 references indexed in Scilit:
- Automatic detection of small lung nodules on CT utilizing a local density maximum algorithmJournal of Applied Clinical Medical Physics, 2003
- Lung Cancer Screening with CT: Mayo Clinic ExperienceRadiology, 2003
- Computerized detection of pulmonary embolism in spiral CT angiography based on volumetric image analysisIEEE Transactions on Medical Imaging, 2002
- Automated detection of pulmonary nodules in helical CT images based on an improved template-matching techniqueIEEE Transactions on Medical Imaging, 2001
- Patient-specific models for lung nodule detection and surveillance in CT imagesIEEE Transactions on Medical Imaging, 2001
- Computer-aided diagnosis in chest radiography: a surveyIEEE Transactions on Medical Imaging, 2001
- How Good Is Intercenter Agreement in the Identification of Embolic Signals in Carotid Artery Disease?Stroke, 1996
- Computerized Detection of Pulmonary Nodules in Computed Tomography ImagesInvestigative Radiology, 1994
- Pulmonary nodules in spiral volumetric and single slice computed tomographyEuropean Journal of Radiology, 1994
- Variables Affecting Pulmonary Nodule Detection with Computed TomographyJournal of Thoracic Imaging, 1993