The look‐up table: A classifier for cell sorters
Open Access
- 1 May 1981
- Vol. 1 (6) , 394-396
- https://doi.org/10.1002/cyto.990010608
Abstract
Currently used classifiers for flow cell sorters are described and the look‐up table is introduced as such. An experimental setup of a look‐up table classifier added to a commercially available cell sorter, under control of a microcomputer is described. The contents of the look‐up table are obtained from contrours of regions of interest. These regions are defined by the user who designates the areas on a video color display of the bivariate flow data. Since most cell clusters form elliptic or ovoid shapes in a multi‐dimensional feature space this method enables a better matching of window shapes to cluster outlines, compared to conventional sort decision circuitry.Keywords
This publication has 3 references indexed in Scilit:
- Interactive display and analysis of data from bivariate flow cytometers.Journal of Histochemistry & Cytochemistry, 1979
- Visualization of multidimensional spectra in flow cytometry.Journal of Histochemistry & Cytochemistry, 1979
- COMPUTER-CONTROLLED MULTIPARAMETER ANALYSIS AND SORTING OF CELLS AND PARTICLESJournal of Histochemistry & Cytochemistry, 1974