Resolution Determination in Field Emission Energy Analyzers
- 1 October 1968
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 39 (10) , 1477-1480
- https://doi.org/10.1063/1.1683139
Abstract
The influence of energy analyzer transmission function on measured field emission energy distributions has been evaluated by folding a Gaussian transmission function into the field emission total energy distribution. The resulting distributions at 0, 4.2, 20.3, 77, and 300 K are plotted as a function of the field‐ and work‐function‐dependent parameter d, and for the special case φ=4.40 eV, F=3.0×107 V/cm. A criterion is established for determining the width of an analyzer transmission function from distributions measured at the above temperatures. It is shown that the slope of the log of the energy distribution is changed by less than 1% when the analyzer width is changed from 0 to 100 mV.Keywords
This publication has 15 references indexed in Scilit:
- Total-Energy Distribution of Field-Emitted Electrons and Single-Plane Work Functions for TungstenPhysical Review B, 1967
- Anomalous Total-Energy Distribution for a Molybdenum Field EmitterPhysical Review Letters, 1967
- Field-Induced Resonance States at a SurfacePhysical Review B, 1967
- Energy Exchanges Attending Field Electron EmissionPhysical Review B, 1966
- ANOMALOUS WORK FUNCTION OF THE TUNGSTEN (110) PLANEApplied Physics Letters, 1966
- Effect of Surface Patch Fields on Field-Emission Work-Function DeterminationsPhysical Review Letters, 1966
- Anomalous Total Energy Distribution for a Tungsten Field EmitterPhysical Review Letters, 1966
- Progress in Field-Emission Work-Function Measurements of Atomically Perfect Crystal PlanesJournal of Applied Physics, 1962
- Theoretical Total-Energy Distribution of Field-Emitted ElectronsPhysical Review B, 1959
- Experimental Measurement of the Total-Energy Distribution of Field-Emitted ElectronsPhysical Review B, 1959