A multicentre prospective randomized controlled trial of induction of labour with an automatic closed‐loop feedback controlled oxytocin infusion system
- 1 November 1985
- journal article
- research article
- Published by Wiley in BJOG: An International Journal of Obstetrics and Gynaecology
- Vol. 92 (11) , 1127-1133
- https://doi.org/10.1111/j.1471-0528.1985.tb03023.x
Abstract
Summary. A randomized prospective controlled trial of an automatic feed‐back controlled oxytocin infusion system (AIS) for the induction of labour at term was carried out in three hospitals. In primiparae, a four‐fold reduction in the total dose of oxytocin infused, for the same induction‐delivery interval, was achieved with the AIS compared with standard labour ward protocols relying on midwife control of oxytocin infusion rate. In multiparae, the total dose of oxytocin infused was almost halved when the AIS was used, while the mean duration of labour was increased by 1.6 h to 7.5 h. There were no differences in the mode of delivery or Apgar scores. It is concluded that the AIS is a safe technique for induction of labour. Its use has demonstrated that term labour can be induced efficiently with a mean oxytocin infusion rate between 2–5 and 3 mU/min (max 8.8 mU/min) and a total oxytocin dose infused of < 5 U.This publication has 15 references indexed in Scilit:
- Normal levels of active contraction area in spontaneous labourBJOG: An International Journal of Obstetrics and Gynaecology, 1984
- Pathogenesis of neonatal hyperbilirubinaemia after induction of labour with oxytocin.BMJ, 1979
- THE EFFECT OF OXYTOCIN IN INDUCED LABOUR ON NEONATAL JAUNDICEBJOG: An International Journal of Obstetrics and Gynaecology, 1979
- THE USE OF CATHETER‐TIP PRESSURE TRANSDUCERS FOR THE MEASUREMENT OF INTRAUTERINE PRESSURE IN LABOURBJOG: An International Journal of Obstetrics and Gynaecology, 1978
- INVESTIGATION OF SOME PROBLEMS ASSOCIATED WITH THE CARDIFF INFUSION PUMPBJOG: An International Journal of Obstetrics and Gynaecology, 1976
- A Critical Assessment of the Induction and Maintenance of Labour Using the Cardiff Infusion System, Mark IIAustralian and New Zealand Journal of Obstetrics and Gynaecology, 1975
- UTERINE ACTIVITY IN lNDUCED LABOURBJOG: An International Journal of Obstetrics and Gynaecology, 1975
- OXYTOCIN‐INDUCED RUPTURE OF THE PRIMIGRAVID UTERUSBJOG: An International Journal of Obstetrics and Gynaecology, 1973
- OXYTOCIC AGENTS AND NEONATAL HYPERBILIRUBINÆMIAThe Lancet, 1972
- AUTOMATIC OXYTOCIN INFUSION EQUIPMENT FOR INDUCTION OF LABOURBJOG: An International Journal of Obstetrics and Gynaecology, 1970