Evaluation of a continuous flow paramagnetic oxygen analyser: the Junkalor Spirolyt.
Open Access
- 31 October 1966
- Vol. 28 (6) , 776-779
- https://doi.org/10.1136/hrt.28.6.776
Abstract
Measurements of O2 uptake and CO2 production made on the Spirolyt paramagnetic analyser and by micro-Sholandermeasurement of expired air were compared at rest and on exercise. Fifty-seven duplicate observations were made on 12 normal subjects at rest and at several work loads. The results indicate satisfactory agreement and show that despite some difficulties in application and Spirolyt can be used for continuous monitoring of O2 uptake and CO2 production at work loads of up to 1200 kpm/min.This publication has 0 references indexed in Scilit: