Analysis of Haemorheological Variables—Methodology and Reference Values
- 18 January 1989
- journal article
- research article
- Published by Uppsala Medical Society in Upsala Journal of Medical Sciences
- Vol. 94 (1) , 81-87
- https://doi.org/10.3109/03009738909179250
Abstract
Routine methods are proposed for the analysis of haemorheologic variables with only one instrument, a Low Shear 30 rotational couette viscometer. The variables are plasma viscosity, whole blood apparent viscosity, erythrocyte aggregation tendency and erythrocyte fluidity. These variables, in addition to erythrocyte volume fraction, were analysed in a population of 83 healthy hospital staff subjects. The following means and standard deviations were found: Erythrocyte volume fraction, women 42±3 %, men 46±2 %; Plasma viscosity 1.31±0.07 mPa-s; Whole blood apparent viscosity, women 4.3±0.4 mPa-s, men 4.8±0.3 mPa-s; Erythrocyte aggregation tendency 1.00±0.10; Erythrocyte fluidity 122±10 Pa−1.s−1.This publication has 4 references indexed in Scilit:
- Blood viscosity - A comparative study on three rotational instrumentsBiorheology, 1985
- Increased blood viscosity in young women using oral contraceptivesAmerican Journal of Obstetrics and Gynecology, 1980
- Altered haemorheology in oral-contraceptive users.BMJ, 1980
- Measurement of blood viscosity [1]12Biorheology, 1974