Transtracheal Doppler
Open Access
- 1 January 1989
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesiology
- Vol. 70 (1) , 134-138
- https://doi.org/10.1097/00000542-198901000-00025
Abstract
An endotracheal tube with a 5 mm diameter ultrasonic transducer incorporated into its distal end was inserted into the tracheas of six anesthetized dogs. The device provides for ultrasound measurement of aortic diameter and blood velocity in the ascending aorta to obtain cardiac outputs. Because measurements are made proximal to the origins of the aortic arch vessels, total cardiac output is obtained. Procedures for measurement of aortic diameter, aortic blood velocity, and cardiac output are discussed. The resulting data were used to empirically determine the angle of incidence of the ultrasound beam that gave the best least squares fit over a range of thermodilution cardiac outputs. For an angle of incidence of 62.3.degree., the relationship between cardiac output determined by thermal dilution and by transtracheal Doppler was transtracheal Doppler cardiac output = -69+ + 1.19 thermal dilution cardiac output. R2 = .82. The results from this preliminary assessment of the feasibility of measuring cardiac output in dogs using transtracheal Doppler suggest that additional studies should be pursued. Additional issues that require investigation include confirmation of constancy of human anatomy and demonstration of lack of tracheal trauma form the Doppler equipped endotracheal tube.Keywords
This publication has 3 references indexed in Scilit:
- Continuous Noninvasive Monitoring of Cardiac Output with Esophageal Doppler Ultrasound during Cardiac SurgeryAnesthesia & Analgesia, 1986
- Noninvasive Doppler determination of cardiac output in man. Clinical validation.Circulation, 1983
- Reliability of the Determination of Cardiac Output in Man by Means of the Fick PrincipleCirculation Research, 1958