A modified single breath method for estimation of cardiac output in humans at rest and during exercise
- 1 April 1987
- journal article
- research article
- Published by Portland Press Ltd. in Clinical Science
- Vol. 72 (4) , 437-441
- https://doi.org/10.1042/cs0720437
Abstract
This study was undertaken to determine the accuracy of a modification of a single breath method for estimation of cardiac output. The technique incorporated a single rebreathing stage followed by a prolonged expiration. Cardiac output was determined from the O2 uptake and the instantaneous changes in O2 and CO2 in the expired gas during the prolonged expiration. The mean values and the random errors (determined from the differences between pairs of estimates) of cardiac outputs in normal subjects at rest and exercise were 5.42 and .+-. 0.60 litres/min (2 Sd, 60 pairs) and 14.1 and .+-. 1.8 litres/min (40 pairs). Larger random errors were obtained in a group of cardiac patients but, except in hypoxic patients, the mean values obtained by the single breath and the direct (Fick) methods were almost identical. We conclude that our modification of the single breath method is simple to use and sufficiently reliable for use in humans both at rest and during steady states of light exercise.This publication has 5 references indexed in Scilit:
- Comparison of cardiac output during exercise by single-breath and CO2-rebreathing methodsJournal of Applied Physiology, 1985
- SINGLE BREATH CARDIAC-OUTPUT - ENHANCED SAMPLING AND ANALYSIS TECHNIQUES1983
- Evaluation using dogs of a method for estimating mixed venous and arterial PCO2 from a single breathJournal of Applied Physiology, 1981
- Evaluation using dogs of a method for estimating cardiac output from a single breathJournal of Applied Physiology, 1981
- Estimation of true venous and arterial PCO2 by gas analysis of a single breath.Journal of Applied Physiology, 1966