Comparability and reproducibility of the results of water loss measurements: a study of 4 evaporimeters
- 11 April 1989
- journal article
- research article
- Published by Wiley in Contact Dermatitis
- Vol. 20 (4) , 241-246
- https://doi.org/10.1111/j.1600-0536.1989.tb03139.x
Abstract
A study was made of the comparability and reproducibility of the results of measurements of water loss, both in vitro and in vivo, using 4 Servo Med Evaporimeters (3 single–probe and 1 double–probe instrument). The optimum time for recording transepidermal water loss (TEWL) after the initial application of the probe to the skin (in vivo), and the best technique for optimizing the accuracy of measuring TEWL were determined. An evaporation device with a constant level of water loss was constructed for in vitro studies. The volar aspect of the right forearm skin of one subject was used in vivo. Both in vitro and in vivo measurements showed that there were some differences between the results of 4 of the 5 probes. The other probe was distinctly out of range. For all probes, the reproducibility of results of successive measurements was high. Stabilization of TEWL values was reached for all probes from 30–45 s after their initial application to the skin. It is recommended that TEWL be recorded for a further 30–s period, after the initial stabilization (45 s), and that this be taken as the true value. The manufacturer's recommended calibration procedure is based only on adjustments for the standard specified humidities and zero water loss. The importance of incorporating an additional calibration procedure which includes adjustments for an actual standard constant water loss is thus strongly stressed.Keywords
This publication has 14 references indexed in Scilit:
- New method to assess the water vapour permeance of wound coveringsBiomaterials, 1988
- Changes in water vapor loss from the skin of metal industry workers monitored during exposure to oils.Scandinavian Journal of Work, Environment & Health, 1986
- The use of transepidermal water loss to measure and predict the irritation response to surfactantsInternational Journal of Cosmetic Science, 1986
- Transepidermal water loss in dry and clinically normal skin in patients with atopic dermatitisActa Dermato-Venereologica, 1985
- Vulnerability of the skin to surfactants in different groups of eczema patients and controls as measured by water vapour lossClinical and Experimental Dermatology, 1985
- Skin Irritancy of Surfactants As Assessed by Water Vapor Loss MeasurementsJournal of Investigative Dermatology, 1984
- Performance of a skin evaporimeterMedical & Biological Engineering & Computing, 1980
- A Method to Evaluate Skin Moisturizers in VivoJournal of Investigative Dermatology, 1978
- TRANSEPIDERMAL WATER LOSS IN PATIENTS WITH DERMATITISBritish Journal of Dermatology, 1969
- Observations on Regeneration of the Skin Barrier**From the Departments of Dermatology, Boston University School of Medicine, Boston, Massachusetts and Medical School, University of Miami, Miami, Florida.This work was supported by grants A-5779, A-5924 and 2G-224 from the National Institutes of Health and by U. S. Army Contract No. DA-49-007-MD-731.Journal of Investigative Dermatology, 1962