Thoracic impedance changes measured via defibrillator pads can monitor ventilation in critically ill patients and during cardiopulmonary resuscitation*
- 1 September 2006
- journal article
- editorial
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 34 (9) , 2399-2405
- https://doi.org/10.1097/01.ccm.0000235666.40378.60
Abstract
Monitoring of ventilation performance during cardiopulmonary resuscitation would be desirable to improve the quality of cardiopulmonary resuscitation. To investigate the potential for measuring ventilation rate and inspiration time, we calculated the correlation in waveform between transthoracic impedance measured via defibrillator pads and tidal volume given by a ventilator. Clinical study. Emergency department of a tertiary care university hospital. A convenience sample of mechanical ventilated patients (n = 32), cardiac arrest patients (n = 20), and patients after restoration of spontaneous circulation (n = 31) older than 18 were eligible. The Heartstart 4000SP defibrillator (Laerdal Medical Cooperation, Stavanger, Norway) with additional capabilities of recording thoracic impedance changes was used. The relationship between impedance change and tidal volume (impedance coefficient) was calculated. The mean (sd) correlations between the impedance waveform and the tidal volume waveform in the patient groups studied were .971 (.027), .969 (.032), and .967 (.035), respectively. The mean (sd) impedance coefficient for all patients in the study was .00194 (.0078) Omega/mL, and the mean (sd) specific (weight-corrected) impedance coefficient was .152 (.048) Omega/kg/mL. The measured thorax impedance change for different tidal volumes (400-1000 mL) was approximately linear. The impedance sensor of a defibrillator is accurate in identifying tidal volumes, when chest compressions are interrupted. This also allows quantifying ventilation rates and inspiration times. However this technology, at its present state, provides only limited practical means for exact tidal volume estimation.Keywords
This publication has 32 references indexed in Scilit:
- Chest Compression Rates During Cardiopulmonary Resuscitation Are SuboptimalCirculation, 2005
- Quality of Cardiopulmonary Resuscitation During In-Hospital Cardiac ArrestJAMA, 2005
- Quality of Cardiopulmonary Resuscitation During Out-of-Hospital Cardiac ArrestJAMA, 2005
- Hyperventilation-Induced Hypotension During Cardiopulmonary ResuscitationCirculation, 2004
- Oxygen delivery and return of spontaneous circulation with ventilation:compression ratio 2:30 versus chest compressions only CPR in pigsResuscitation, 2004
- Improving CPR performance using an audible feedback system suitable for incorporation into an automated external defibrillatorResuscitation, 2003
- Expanding automatic external defibrillators to include automated detection of cardiac, respiratory, and cardiorespiratory arrestCritical Care Medicine, 2002
- An automated voice advisory manikin system for training in basic life support without an instructor. A novel approach to CPR trainingResuscitation, 2001
- Tidal volumes which are perceived to be adequate for resuscitationResuscitation, 1996
- Central and regional blood flow during hyperventilationActa Anaesthesiologica Scandinavica, 1994