Frameless stereotaxy with real-time tracking of patient head movement and retrospective patient—image registration
- 1 August 1996
- journal article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 85 (2) , 287-292
- https://doi.org/10.3171/jns.1996.85.2.0287
Abstract
✓ The accuracy of a novel frameless stereotactic system was determined during 10 surgeries performed to resect brain tumors. An array of three charge-coupled device cameras tracked the locations of infrared light-emitting diodes on a hand-held stylus and on a reference frame attached to the patient's skull with a single bone screw. Patient—image registration was achieved retrospectively by digitizing randomly chosen scalp points with the system and fitting them to a scalp surface model derived from magnetic resonance (MR) images. The reference frame enabled continual correction for patient head movements so that registration was maintained even when the patient's head was not immobilized in a surgical clamp. The location of the stylus was displayed in real-time on cross-sectional and three-dimensional MR images of the head; this information was used to predict the locations of small intracranial lesions. The average distance (and standard deviation) between the actual position of the mass and its stereotactically predicted location was 4.8 ± 3.5 mm. The authors conclude that frameless stereotaxy can be used for accurate localization of intracranial masses without resorting to using fiducial markers during presurgical imaging and without immobilizing the patient's head during surgery.Keywords
This publication has 13 references indexed in Scilit:
- Interactive 3D patient — image registrationPublished by Springer Nature ,2005
- Craniotomy simulation and guidance using a stereo video based tracking system (VISLAN)Published by SPIE-Intl Soc Optical Eng ,1994
- A frameless, armless navigational system for computer-assisted neurosurgeryJournal of Neurosurgery, 1991
- Three-dimensional magnetic resonance images of the brain: application to neurosurgical planningJournal of Neurosurgery, 1990
- The brain: integrated three-dimensional display of MR and PET images.Radiology, 1989
- Surface of the brain: three-dimensional MR images created with volume rendering.Radiology, 1989
- Accurate Three-Dimensional Registration of CT, PET, and/or MR Images of the BrainJournal of Computer Assisted Tomography, 1989
- Volumetric Stereotactic Surgical Resection of Intra-axial Brain Mass LesionsMayo Clinic Proceedings, 1988
- A new imaging method for intraoperative therapy control in skull-base surgeryNeurosurgical Review, 1988
- A frameless stereotaxic integration of computerized tomographic imaging and the operating microscopeJournal of Neurosurgery, 1986