Pulsatile Flow and Simple Flow Control Method During Weaning Period in Centrifugal Pump: Toward More Expanded Usage in Open Heart Surgery
- 1 September 1994
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 18 (9) , 643-649
- https://doi.org/10.1111/j.1525-1594.1994.tb03393.x
Abstract
To expand the usage of the centrifugal pump (CP) in open heart surgery, we performed two studies. In the first, we evaluated pulsatile flow in the CP. In vitro pump performance of the Terumo Capiox pump (TCP) and the Sarns Delphin pump (SDP) and increase of free hemoglobin (mg/dl) after driving 6 h were investigated using bovine blood. A roller pump (RP) was used as a comparison. Equally effective pulsatile flow was obtained in both CPs. Hemolysis was less severe in TCP (120 mg/dl) than SDP (210 mg/dl) and RP (320 mg/dl). In the second study, we evaluated a simple flow control method. Flow rate was easily controlled with step-wise clamping of 3-pronged tubing (Triple-flow) without changing rotational speed, regardless of afterload. Fluctuation of flow was much less with this method than with the rotational speed change method. The use of pulsatile flow of TCP, with its minimum increase of hemolysis and the easier flow control method during the weaning process, may expand the usage of CP in open heart surgery.Keywords
This publication has 4 references indexed in Scilit:
- Clinical Experience of Assisted Circulation with a Centrifugal Pump at Tokyo Women's Medical CollegeArtificial Organs, 1993
- A Straight Path Centrifugal Blood Pump Concept in the Capiox Centrifugal PumpArtificial Organs, 1993
- Development of the Terumo Capiox Centrifugal Pump and Its Clinical Application to Open Heart Surgery: A Comparative Study with the Roller PumpArtificial Organs, 1993
- Comparative clinical study of pulsatile and non-pulsatile perfusion in 350 consecutive patients.Thorax, 1982