High-sensitivity inductive magnetometer for pulsed magnetic fields
- 1 January 1999
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 70 (1) , 82-84
- https://doi.org/10.1063/1.1149544
Abstract
A magnetometer is described for measurements in pulsed magnetic fields up to 43 T. The compensation system allows removal of pickups produced both by the longitudinal and transverse components of the pulsed field. The compensation level achieves 10 −6 (of the signal from the working coil with an inner diameter 1.4 or 1.75 mm) for the longitudinal and 10 −4 for the transverse components. Sensitivity of the magnetometer is about 10 −4 emu for the fields up to 10 T and decreases down to 10 −3 emu at the pulse amplitude of 35 T. A deviation of the compensation is below 2×10 −4 for the temperature increase from 77 up to 300 K and below 5×10 −5 after the sample replacement. The sample temperature is controlled by a fast-acting temperature control system in the range from 6 up to 300 K with accuracy of 0.3–0.05 K.Keywords
This publication has 5 references indexed in Scilit:
- A probe for magnetization measurements of thin superconducting films in pulsed high magnetic fieldsReview of Scientific Instruments, 1995
- Magnetization measurements in very high pulsed fields produced by a single-turn coil systemPhysica B: Condensed Matter, 1989
- Sensitive magnetometer available to 10-4 EMU using pulsed magnetPhysica B: Condensed Matter, 1989
- Magnetic measurements of high Tc superconductors in megagauss fieldsPhysica B: Condensed Matter, 1989
- Superconductivity in Yba2Cu3O7–x in a 100 tesla magnetic fieldNature, 1988