A battery state-of-charge indicator for electric wheelchairs
- 1 October 1992
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Industrial Electronics
- Vol. 39 (5) , 398-409
- https://doi.org/10.1109/41.161471
Abstract
Deep-discharge-type lead-acid batteries used in most electric wheelchairs require acurrate battery state-of-charge indication to prevent stranding and to provide economical operation of the wheelchair. A monitoring technique combining the open-circuit voltage and the coulometric measurements had been previously implemented on a microcomputer-based circuit. This adaptive monitoring technique enables the monitor to adjust to different battery sizes as well as the aging process. Several improvements are reported. A technique has been developed to enhance the acurracy and reduce the required rest period of the open-circuit voltage measurement. The open-circuit voltage recovery curve is approximated by two asymptotes on a semilog scale. The open-circuit voltage is then extrapolated from the slope of the first asymptote before it fully stabilizes. The accuracy of the monitor has been verified in field tests, and comparison with a commercial battery monitor shows it to be superior in several respects.<>Keywords
This publication has 0 references indexed in Scilit: