An application of GafChromic MD-55 film for 67.5 MeV clinical proton beam dosimetry
- 6 October 1999
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 44 (11) , 2735-2745
- https://doi.org/10.1088/0031-9155/44/11/304
Abstract
The purpose of this study is to explore the use of GafChromic MD-55 (RC) film for 67.5 MeV clinical proton beam dosimetry at the Crocker Nuclear Laboratory, University of California, Davis. Several strips of RC film 6 cm × 6 cm in dimension were irradiated at a depth of 18.2 mm corresponding to the middle of a 24 mm spread-out Bragg peak (SOBP). The films were irradiated to a proton dose in the range of 0.5 Gy to 100 Gy. The beam profiles were also measured at the middle of the 24 mm SOBP. The Bragg peak was measured by using a wedge shaped phantom made of Lucite. The Bragg peak measured with RC film was compared with diode and ionization chamber measurements. After background subtraction, the calibration of the dose response of RC film showed, to a maximum deviation of 10%, a linear increase of optical density (OD) with dose from 0.5 to 100 Gy. The uniformity of OD over a single sheet of film showed a variation of ±6%. The distal-fall off between 90% and 20% measured with GafChromic film for the Bragg peak was 1.3 mm as compared to 1.1 mm for a diode measurement and 1.4 mm for an ionization chamber measurement. The FWHM of the Bragg peak was 7.5 mm when measured with GafChromic film, 5.3 mm when measured with a diode and 8.1 mm as measured by an ionization chamber. The peak/plateau ratio with GafChromic film was 3.3 as compared to 3.7 with a diode and 3.2 with an ionization chamber. In conclusion, GafChromic MD-55 film may be a useful and convenient detector for dose measurement and quality assurance programmes of proton beams.Keywords
This publication has 22 references indexed in Scilit:
- Comments on “Quantitative evaluation of radiochromic film response for two-dimensional dosimetry” [Med. Phys. 24 , 223-231 (1997)]Medical Physics, 1998
- Fast 2D phantom dosimetry for scanning proton beamsMedical Physics, 1998
- GafChromic MD‐55: Investigated as a precision dosimeterMedical Physics, 1997
- New UCSF proton ocular beam facility at the Crocker Nuclear Laboratory Cyclotron (UC Davis)Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1996
- The use of a radiochromic detector for the determination of stereotactic radiosurgery dose characteristicsa)Medical Physics, 1994
- NMR relaxation enhancement in gels polymerized and cross-linked by ionizing radiation: A new approach to 3D dosimetry by MRIMagnetic Resonance Imaging, 1993
- Sensitometry of the response of a new radiochromic film dosimeter to gamma radiation and electron beamsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1991
- GafchromicTM dosimetry media: A new high dose, thin film routine dosimeter and dose mapping toolInternational Journal of Radiation Applications and Instrumentation. Part C. Radiation Physics and Chemistry, 1990
- The response of plastic dosimeters to gamma rays and electrons at high absorbed dose ratesRadiation Physics and Chemistry (1977), 1979
- Radiochromic plastic films for accurate measurement of radiation absorbed dose and dose distributionsRadiation Physics and Chemistry (1977), 1977