Interference Due to Haemolysis in Routine Photometric Analysis—A Survey

Abstract
To assess the affect of haemolysis on routine photometric methods used by participants in the Irish External Quality Assessment Scheme for Clinical Biochemistry (1EQAS) a sample was distributed with a supplied diluent containing haemoglobin. A sample of the same batch had previously been circulated as a normal distribution. Many analytes/methods were affected by the haemolysis. The range of values reported for each analyte was as follows. Albumin 34–48 g/L (baseline 37 g/L). Calcium 2.16–2.68 mmol/L (baseline 2.34 mmol/L). Urate 90–470 μmol/L (baseline 195 μmol/L). Creatinine 55–161 μmol/L (baseline 140 μmol/L). Glucose 4.1–5.6 mmol/L (baseline 4.8 mmol/L). Bilirubin 14–60 μmol/L (baseline 36 μmol/L). Urea 3.4–7.3 mmol/L (baseline 3.8 mmol/L). Iron 17.9–55.0 μmol/L (baseline 26.8 μmol/L). Inorganic phosphorus 0.22–2.41 mmol/L (baseline 1.59 mmol/L). Cholesterol 3.0–5.9 mmol/L (baseline 4.1 mmol/L). In some situations the use of an appropriate blank may solve the problem, but it must be remembered that interference caused by haemoglobin may be more complex than merely spectral.

This publication has 6 references indexed in Scilit: