The development of equipment for the technical assessment of respiratory motion induced artefacts in MRI
- 1 February 1993
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 66 (782) , 132-139
- https://doi.org/10.1259/0007-1285-66-782-132
Abstract
A device and technique to study the effects of respiratory motion on the quality of magnetic resonance images is proposed. The construction of the device enables a variety of test objects to be mounted and used in the evaluation of imaging parameters that may be affected by motion. The equipment is constructed of cast acrylic and the movement is actuated and controlled pneumatically thus ensuring that there are no interactions with the magnetic field and radiofrequency detection system to cause further image artefacts. Separate studies have been performed, using ultrasound, to assess the degree and rate of movement of organs owing to respiration in order to derive the motion parameters for the apparatus. Preliminary results indicate that the technique produces motion induced artefacts simulating those which are the result of the effects of respiration.Keywords
This publication has 15 references indexed in Scilit:
- Breath-hold T2-weighted sequences of the liver: A comparison of four techniques at 1.0 and 1.5 TMagnetic Resonance Imaging, 1992
- Respiratory kinematics of the upper abdominal organs: A quantitative studyMagnetic Resonance in Medicine, 1992
- Effect of bulk tissue motion on quantitative perfusion and diffusion magnetic resonance imagingMagnetic Resonance in Medicine, 1991
- Spatial localization for motion-rejected nmr imaging: slo-motionMagnetic Resonance in Medicine, 1990
- Compensation for effects of linear motion in MR imagingMagnetic Resonance in Medicine, 1989
- Analysis of imaging axes significance in motion artifact suppression technique (MAST™): MRI of turbulent flow and motionMagnetic Resonance Imaging, 1989
- A method for reducing motion induced errors in T 2 -weighted magnetic resonance imagingMedical Physics, 1987
- MR artifacts: a reviewAmerican Journal of Roentgenology, 1986
- Reducing motion artifacts in two-dimensional Fourier transform imagingMagnetic Resonance Imaging, 1986
- Relative contributions of rib cage and abdomen to breathing in normal subjectsJournal of Applied Physiology, 1975