Variations in Mechanical Sampling: Their Causes and Effects
Open Access
- 1 January 1975
- journal article
- research article
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 12 (1-6) , 145-150
- https://doi.org/10.1177/000456327501200135
Abstract
Three characteristics of a sample of AutoAnalyzer Sampler II modules have been measured : eccentricity of the cam, eccentricity of the spindle driving the cam, and the lengths of the chords of the cam lobes. Imprecision in timing may make a significant contribution to analytical error.Keywords
This publication has 4 references indexed in Scilit:
- Curve regeneration in continuous flow analysisClinica Chimica Acta; International Journal of Clinical Chemistry, 1971
- Practical considerations in kinetics of continuous flow analysisClinica Chimica Acta; International Journal of Clinical Chemistry, 1970
- Variations in Sampling Cam Specifications for Continuous-Flow Automated SystemsClinical Chemistry, 1970