CO2 lasers: Beam patterns in relation to surgical use
- 1 January 1983
- journal article
- research article
- Published by Wiley in Lasers in Surgery and Medicine
- Vol. 2 (4) , 331-341
- https://doi.org/10.1002/lsm.1900020406
Abstract
According to surgeons operating with a variety of CO2 lasers available at the National Cancer Institute of Milan (Coherent, Sharplan, Valfivre), these lasers have different cutting and coagulation properties. To identify what physical parameters might corroborate the subjective impression of the surgeons, a comparative study of the crater forms in perspex samples was performed. Perspex was chosen for its thermal properties (in fact, its thermal conductivity and diffusivity are similar to those of organic tissue) and because it allowed good visualization and measurement of crater characteristics. Depth of penetration, crater diameter, and extension of thermal damage were measured against power, focal length, and exposure time for each CO2 laser model. These results can be used as an index of behaviour of different surgical lasers. It appears that for fully characterizing the interaction of surgical lasers with the sample, it is necessary to specify either power, focal length, exposure time, or beam mode.Keywords
This publication has 4 references indexed in Scilit:
- Laser Related Heat Effects on Blood VesselsPublished by Springer Nature ,1980
- Transmissive Optics for High Power CO2 Lasers: Practical ConsiderationsOptical Engineering, 1978
- USE OF THE CARBON DIOXIDE LASER FOR LARGE EXCISIONS WITH MINIMAL BLOOD LOSSPlastic and Reconstructive Surgery, 1977
- Properties of Low Order Transverse Modes in Argon Ion LasersApplied Optics, 1969