Axonal stimulation for end-plate jitter studies.
Open Access
- 1 June 1986
- journal article
- research article
- Published by BMJ in Journal of Neurology, Neurosurgery & Psychiatry
- Vol. 49 (6) , 677-685
- https://doi.org/10.1136/jnnp.49.6.677
Abstract
This single fibre EMG study compares the standard method of neuromuscular jitter measurement in voluntarily activated muscle to that by intramuscular electrical stimulation of motor axons in a group of normal subjects. The latter method avoids the interdischarge interval-dependent jitter, as well as a possible failure to recognise split muscle fibres. The mean MCD on axonal stimulation was only 5.2 microseconds less than in the voluntary activation study and was thus 8% more than theoretically expected for single motor end plates. The difference could be due to an axonal jitter and some other factors. Axonal stimulation has proved to be a relatively easy and reliable method for routine estimation of neuromuscular jitter, provided that the resolution of time measurement is better than 2 microseconds, so that low jitter due to occasional direct muscle fibre stimulation is not mistaken for a normal reading. The upper normal limits for the extensor digitorum communis muscle suggested by the present study are 40 microseconds (individual muscle fibres) and 25 microseconds (mean of 30 muscle fibres).This publication has 4 references indexed in Scilit:
- Causes of the increased fiber density in muscular dystrophies studied with single fiber EMG during electrical stimulationMuscle & Nerve, 1985
- Responses to electrical stimulation of denervated human muscle fibres recorded with single fibre EMG.Journal of Neurology, Neurosurgery & Psychiatry, 1983
- Jittermeter—A microcomputer-based system for single fibre electromyographyInternational Journal of Bio-Medical Computing, 1979
- Demonstration of axon reflexes in human motor nerve fibresJournal of Neurology, Neurosurgery & Psychiatry, 1970