Conundrums with FlowPRA™ beads
- 1 October 2002
- journal article
- research article
- Published by Wiley in Clinical Transplantation
- Vol. 16 (s7) , 24-29
- https://doi.org/10.1034/j.1399-0012.16.s7.3.x
Abstract
In 1969, a study by Patel and Terasaki persuaded the renal transplant community that a pre-transplant cross-match should always be performed between donor and recipient to detect HLA antibodies and prevent hyperacute allograft rejection. Although the role of the cross-match among nonsensitized patients is controversial, its importance among sensitized recipients is undeniable. Over the past 30 years, more sensitive techniques, such as the flow cytometric cross-match (FCXM), were developed to identify low levels of antibodies undetectable by other approaches. The clinical relevance of a positive FCXM, however, has been hotly disputed, with some investigators maintaining that the FCXM is ‘too sensitive’ and rules out acceptable donor–recipient combinations. An alternative explanation is that the FCXM is non-specific, and, at least in certain situations, identifies non-HLA antibodies that are clinically irrelevant. Recently, a solid phase immunoassay utilizing purified HLA Class I or Class II molecules bound to microparticles (FlowPRATM) was developed. Ideally, use of the FlowPRATM for the identification of HLA antibodies in recipient sera would help ascertain whether a positive FCXM with donor cells was truly the result of an HLA-specific antibody. As shown here, this may not always be true. In this study, two unexpected serum patterns were observed. Pattern 1: FlowPRATM beads were positive (with an associated HLA Class I specificity) and the FCXM with cells expressing the HLA antigen(s) to which the antibody was directed, was negative. Sequence analysis of the HLA antigens reactive with this unexpected antibody suggests that the epitope recognized resides on the floor of the groove, a site generally not expected to generate antibody activity. Pattern 2: FlowPRATM beads were negative yet the FCXM was T and B cell positive. Further analysis of the FlowPRATM negative/FCXM positive sera using a flow cytometric cell-based panel reactive antibody (PRA) approach revealed those sera to have specific anti-HLA Class I activity. We suspect that both types of antibodies described above have clinical relevance. Thus, a negative or positive FCXM (when the FlowPRATM against donor antigens is positive or negative, respectively) is not always a straightforward interpretation.Keywords
This publication has 25 references indexed in Scilit:
- FLOW CYTOMETRY-DETECTED IgG IS NOT A CONTRAINDICATION TO RENAL TRANSPLANTATIONTransplantation, 1999
- FLOW CYTOMETRY CROSS-MATCHTransplantation, 1997
- NO ADVANTAGE OF FLOW CYTOMETRY CROSSMATCH OVER COMPLEMENT-DEPENDENT CYTOTOXICITY IN IMMUNOLOGICALLY WELL-DOCUMENTED RENAL ALLOGRAFT RECIPIENTSTransplantation, 1996
- IgG ANTIBODIES AGAINST AN HLA ANTIGEN ARE ASSOCIATED WITH ACTIVATED CYTOTOXIC T CELLS AGAINST THIS ANTIGEN, IgM ARE NOTTransplantation, 1994
- Epitope map of the HLA-B7 CREG using affinity-purified human alloantibody probesHuman Immunology, 1990
- Topographic map of the HLA-A2 CREG epitopes using human alloantibody probesHuman Immunology, 1990
- THE FLOW CYTOMETRIC CROSSMATCH AND EARLY RENAL TRANSPLANT LOSSTransplantation, 1990
- THE FLOW CYTOMETRIC CROSSMATCHTransplantation, 1989
- IMMUNOGLOBULIN CLASS AND SPECIFICITY OF ANTIBODIES CAUSING POSITIVE T CELL CROSSMATCHESTransplantation, 1986
- HYPERACUTE REJECTION OF A RENAL ALLOGRAFT IN THE PRESENCE OF ANTI-HLA-Cw5 ANTIBODYTransplantation, 1986