Abstract
We describe a rapid, kinetic, fixed-time method for determining serum total cholesterol by use of cholesterol esterase, cholesterol oxidase, and the indicator reaction with peroxidase, 4-aminophenazone, and phenol. On addition of the competitive inhibitor 3,4-dichlorophenol the Michaelis constant of cholesterol oxidase is apparently increased, which extends the linear relation between absorbance change and cholesterol concentration to 20.7 to 25.9 mmol/L, depending on the analyzer being used. For calibration, a single standard is used. Total analysis time is in the range of 80 to 210 s. Incubation temperature is 25 degrees C or 37 degrees C. The single-reagent procedure has been adapted to three different centrifugal analyzers and to the Eppendorf ACP 5040 analyzer. It yields precise and accurate results and is insensitive to potential interferences.

This publication has 0 references indexed in Scilit: