Color and conventional image‐directed Doppler ultrasonography: Accuracy and sources of error in quantitative blood flow measurements
- 1 March 1992
- journal article
- research article
- Published by Wiley in Journal of Clinical Ultrasound
- Vol. 20 (3) , 187-193
- https://doi.org/10.1002/jcu.1870200305
Abstract
Accuracy of two systems—conventional (DRF 400, Diasonics) and colorcoded (Angiodynograph, Quantum/Phillips) image‐directed Doppler ultrasonography—was investigated using an in vitro model that generated both monophasic and triphasic pulsatile flow patterns. Estimated and actual blood volume flow rates showed good correlations, but the sampling with a hand‐held transducer led to wide variations in measurement error for the conventional (−69.2% to 50%) and the colorcoded (−79.3% to 265.7%) systems. By performing multiple measurements, one could improve accuracy considering only the maximal values of a series instead of the mean values. Accuracy was impaired by interposed muscular or fatty tissue due to false low time‐average velocity measurements caused by a loss of Doppler signal. Comparison of both systems revealed significant differences between pulsatility index values (p < 0.001), blood flow velocities (p < 0.001), and blood volume flow rates (p < 0.05 for program flow, p < 0.001 for manual and automatic flow program of the color‐coded system).Keywords
This publication has 20 references indexed in Scilit:
- BLOOD VELOCITY WAVEFORMS AND PLACENTAL VASCULAR FORMATIONThe Lancet, 1988
- DOPPLER UTEROPLACENTAL WAVEFORMS IN PREGNANCY-INDUCED HYPERTENSION: A RE-APPRAISALThe Lancet, 1988
- References ranges and sources of variation for indices of pulsed Doppler flow velocity waveforms from the uteroplacental and fetal circulationBJOG: An International Journal of Obstetrics and Gynaecology, 1988
- Duplex scanning of the internal carotid artery: An assessment of cerebral blood flowBritish Journal of Surgery, 1987
- Vertebrobasilar insufficiency: Evaluation by quantitative duplex flow measurementsJournal of Vascular Surgery, 1987
- Clinical application of an ultrasonic duplex system in the quantitative measurement of portal blood flowJournal of Clinical Ultrasound, 1986
- Measurement of mesenteric blood flow by duplex scanningJournal of Vascular Surgery, 1986
- A combined ultrasonic linear array scanner and pulsed Doppler velocimeter for the estimation of blood flow in the foetus and adult abdomen-I: Technical aspectsUltrasound in Medicine & Biology, 1985
- The clinician and the vascular laboratory in the diagnosis of aortoiliac stenosisBritish Journal of Surgery, 1984
- Comparative study of four methods for quantifying doppler ultrasound waveforms from the femoral arteryUltrasound in Medicine & Biology, 1984