Ethylene glycol—water mixtures have been found suitable for use in ultrasound test objects. A plot of the required percentage of ethylene glycol with room temperature for an acoustic velocity of 1540 m/sec is presented to aid the clinical user. The velocity measurements performed are shown to be dominated by near-field transient diffraction effects. A 30 percent ethylene glycol—water mixture has been found to have a constant acoustic velocity of 1638 ± 3.5 m/sec over a wide temperature range (15.6°C–38.4°C). This mixture is suitable for a test object velocity standard. Proper use of the A.I.U.M. 100-mm test object is discussed.Keywords
This publication has 5 references indexed in Scilit:
- Measurement of spatial distribution of refractive index in tissues by ultrasonic computer assisted tomographyUltrasound in Medicine & Biology, 1978
- Changes in Velocity of an Elastic Pulse owing to Geometrical DiffractionThe Journal of the Acoustical Society of America, 1966
- Diffraction Effects in the Ultrasonic Field of a Piston Source and Their Importance in the Accurate Measurement of AttenuationThe Journal of the Acoustical Society of America, 1956
- Ultrasonic Velocity in Ethanol-Water MixturesThe Journal of the Acoustical Society of America, 1947
- Temperature Coefficient of Ultrasonic Velocity in SolutionsThe Journal of the Acoustical Society of America, 1947