PATIENT EFFECTIVE RADIATION DOSE AND ASSOCIATED RISK FROM TRANSMISSION SCANS USING 153GD LINE SOURCES IN CARDIAC SPECT STUDIES
- 1 July 2002
- journal article
- research article
- Published by Wolters Kluwer Health in Health Physics
- Vol. 83 (1) , 66-74
- https://doi.org/10.1097/00004032-200207000-00007
Abstract
The aim of the present study was to determine the contribution of transmission measurements acquisition to total patient effective dose from cardiac SPECT studies. A dual-head L-shaped gamma camera equipped with a transmission scan acquisition system based on two 153Gd line sources was used to simulate transmission measurements acquisition exposure on an anthropomorphic phantom. Thermoluminescence dosimeters were used to directly monitor the dose to 550 measuring points in the phantom. The effective dose and associated risk from transmission scans acquisition were estimated and compared to those associated to the radiopharmaceutical injected. The maximum effective dose from a typical transmission measurements acquisition was 1.3 μSv and 1.9 μSv for male and female patients, respectively. The contribution of the typical transmission scans acquisition to total patient radiation risk from a cardiac SPECT study is less than 10−3. Thus, radiation exposure may not be considered as a limiting factor for the clinical application of attenuation correction methods based on transmission measurements in cardiac SPECT.Keywords
This publication has 18 references indexed in Scilit:
- Image-correction techniques in SPECTComputerized Medical Imaging and Graphics, 2001
- Attenuation corrected cardiac perfusion SPECTCurrent Opinion in Cardiology, 2000
- Estimation of Fetal Radiation Dose from Computed Tomography Scanning in Late PregnancyInvestigative Radiology, 2000
- Use of asymmetric fan‐beam transmission computed tomography for attenuation correction of cardiac SPECT imagingMedical Physics, 2000
- Attenuation and scatter compensation in myocardial perfusion SPECTSeminars in Nuclear Medicine, 1999
- Dosimetry of transmission measurements in nuclear medicine: a study using anthropomorphic phantoms and thermoluminescent dosimetersEuropean Journal of Nuclear Medicine and Molecular Imaging, 1998
- Transmission scanning in emission tomographyEuropean Journal of Nuclear Medicine and Molecular Imaging, 1998
- Maximum Embryo Absorbed Dose from Intravenous Urography: Interhospital VariationsRadiation Protection Dosimetry, 1997
- Accidental Embryo Irradiation During Barium Enema ExaminationsInvestigative Radiology, 1996
- The Relationship of Effective Dose to Personnel and Monitor Reading for Simulated Fluoroscopic Irradiation ConditionsHealth Physics, 1993