The small-animal radiation research platform (SARRP): dosimetry of a focused lens system
- 26 April 2007
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 52 (10) , 2729-2740
- https://doi.org/10.1088/0031-9155/52/10/007
Abstract
A small animal radiation platform equipped with on-board cone-beam CT and conformal irradiation capabilities is being constructed for translational research. To achieve highly localized dose delivery, an x-ray lens is used to focus the broad beam from a 225 kVp x-ray tube down to a beam with a full width half maximum (FWHM) of approximately 1.5 mm in the energy range 40-80 keV. Here, we report on the dosimetric characteristics of the focused beam from the x-ray lens subsystem for high-resolution dose delivery. Using the metric of the average dose within a 1.5 mm diameter area, the dose rates at a source-to-surface distance (SSD) of 34 cm are 259 and 172 cGy min(-1) at 6 mm and 2 cm depths, respectively, with an estimated uncertainty of +/-5%. The per cent depth dose is approximately 56% at 2 cm depth for a beam at 34 cm SSD.Keywords
This publication has 10 references indexed in Scilit:
- A preliminary investigation of cell growth after irradiation using a modulated x-ray intensity patternPhysics in Medicine & Biology, 2006
- Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energiesMedical Physics, 2005
- Weak energy dependence of EBT gafchromic film dose response in the 50kVp–10MVp X-ray rangeApplied Radiation and Isotopes, 2005
- Precise radiochromic film dosimetry using a flat-bed document scannerMedical Physics, 2005
- In vitroresponse of tumour cells to non-uniform irradiationPhysics in Medicine & Biology, 2005
- Normal tissue radiobiology: from the laboratory to the clinicInternational Journal of Radiation Oncology*Biology*Physics, 2001
- In-field and out-of-field effects in partial volume lung irradiation in rodents: Possible correlation between early dna damage and functional endpointsInternational Journal of Radiation Oncology*Biology*Physics, 2000
- Spatial heterogeneity of the volume effect for radiation pneumonitis in mouse lungInternational Journal of Radiation Oncology*Biology*Physics, 1997
- How water equivalent are water‐equivalent solid materials for output calibration of photon and electron beams?Medical Physics, 1995
- Low-energy x-ray interaction coefficients: Photoabsorption, scattering, and reflectionAtomic Data and Nuclear Data Tables, 1982