Polyethylene wear in uncemented knee replacements
- 1 January 1992
- journal article
- Published by British Editorial Society of Bone & Joint Surgery in The Journal of Bone and Joint Surgery. British volume
- Vol. 74-B (1) , 18-22
- https://doi.org/10.1302/0301-620x.74b1.1732256
Abstract
Isolated wear of the polyethylene tibial component led to failure in five of a series of 108 uncemented porous-coated knee replacements. The clinical features included pain, effusion and instability with progressive varus deformity. In all cases there was extensive wear on the medial side of the polyethylene surface of the prosthesis. The mechanism of such wear is complex, being due in part to the unconstrained nature of the joint and the incongruity of its surfaces. Other design characteristics may have contributed.Keywords
This publication has 0 references indexed in Scilit: