Dosimetric verification of inverse planned step and shoot multileaf collimator fields from a commercial treatment planning system
Open Access
- 1 March 2002
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 3 (2) , 97-109
- https://doi.org/10.1120/jacmp.v3i2.2580
Abstract
An inverse treatment planning (ITP) module on a commercial treatment planning system (TPS) (Helax AB, Uppsala, Sweden) is being used for an in‐house clinical trial for treatment of nasopharyngeal cancer with contralateral parotid sparing. Intensity modulated radiation therapy (IMRT) fields are delivered by step and shoot multileaf collimator (MLC) with a DMLC enabled Varian 2300 CD (Varian Associates, Palo Alto, CA). A series of testing procedures have been devised to quantify the modeling and delivery accuracy of routine clinical inverse planned IMRT using Helax TMS and the Varian step and shoot MLC delivery option. Testing was done on specific aspects of the TPS modeling germane to DMLC. Measured relative dose factors (head scatter plus phantom scatter) for small MLC fields, normalized to a 10×10cm2 non‐MLC field, were found to differ by 2–3% from the TPS values for the smallest of the fields tested. Relative distributions for small off axis fields were found to be in good agreement. A process for the routine clinical verification of IMRT fields has been implemented. Each IMRT field in an inverse plan is imported into a flat water tank plan and a “beam's eye view” (BEV) dose distribution is generated. This is compared to the corresponding measured BEV dose distribution. The IMRT verification process has also been performed using an anthropomorphic phantom. Large clinical fields (i.e., greater than 14.5 cm in the leaf direction) caused difficulties due to a vendor specific machine restriction, and several techniques for dealing with these were examined. These techniques were (i) the use of static stepping of closed junctions, (ii) the use of two separate IMRT fields for a given gantry angle, and (iii) restricting the overall maximum field size used. The overall process has allowed implementation of an in‐house protocol for IMRT use on an initial clinical site. Results of the verification measurements for the first ten patients treated at this center reveal an average maximum dose per IMRT field delivered of 71.0 cGy, with a mean local deviation from the planned dose of – 1.2 cGy, and a standard deviation of 2.4 cGy.Keywords
Funding Information
- Helax AB (MDS Nordion)
This publication has 26 references indexed in Scilit:
- Variation of sensitometric curves of radiographic films in high energy photon beamsMedical Physics, 2001
- Dose, volume, and function relationships in parotid salivary glands following conformal and intensity-modulated irradiation of head and neck cancerInternational Journal of Radiation Oncology*Biology*Physics, 1999
- Theoretical considerations of monitor unit calculations for intensity modulated beam treatment planningMedical Physics, 1999
- A software tool for the quantitative evaluation of 3D dose calculation algorithmsMedical Physics, 1998
- Dynamic multileaf collimation without `tongue-and-groove' underdosage effectsPhysics in Medicine & Biology, 1996
- Parotid gland sparing in patients undergoing bilateral head and neck irradiation: Techniques and early resultsInternational Journal of Radiation Oncology*Biology*Physics, 1996
- Dosimetric verification of intensity‐modulated fieldsMedical Physics, 1996
- Advances in radiation therapy for head and neck cancerCurrent Opinion in Oncology, 1995
- Analytic modeling of photon scatter from flattening filters in photon therapy beamsMedical Physics, 1994
- Does improved depth dose characteristics and treatment planning correlate with a gain in therapeutic results? evidence from past clinical experience using conventional radiation sourcesInternational Journal of Radiation Oncology*Biology*Physics, 1992