Microbial adherence to and disinfection of laryngoscopes used in office practice
Open Access
- 1 May 1993
- journal article
- Published by Wiley in The Laryngoscope
- Vol. 103 (5) , 503-508
- https://doi.org/10.1288/00005537-199305000-00004
Abstract
This study reports the appropriate disinfection methods for flexible and rigid laryngoscopes when used in an outpatient setting. This investigation presents evidence that an appreciable number of laryngoscopes are contaminated during use. A brief tap water rinse serves to eliminate gross soil of host origin known to interfere with disinfectant efficacy. Disinfection with 3.2% glutaraldehyde (Cidexplus) for 10 minutes effectively eliminated microbial contaminants. Since potentially harmful microorganisms were included among the more than 100 isolates during our baseline experiments, the need to eliminate bacteria, viruses, fungi, and protozoa from laryngoscopes is obvious and can be attained readily.Keywords
This publication has 8 references indexed in Scilit:
- Manual of Clinical MicrobiologyRevista do Instituto de Medicina Tropical de São Paulo, 1991
- Elimination of high titre HIV from fibreoptic endoscopes.Gut, 1990
- APIC guideline for selection and use of disinfectantsAmerican Journal of Infection Control, 1990
- CONTAMINATION OF ENDOSCOPES USED IN AIDS PATIENTSThe Lancet, 1989
- Bronchopulmonary Cross-Colonization and Infection Related to Mycobacterial Contamination of Suction Valves of BronchoscopesThe Journal of Infectious Diseases, 1989
- Evaluation of three disinfectants after in-use stressJournal of Hospital Infection, 1988
- Endoscopy and infection.Gut, 1983
- Endoscopic transmission of hepatitis B virus.Gut, 1983