A More Sensitive Automated Method for Determination of Ornithine Carbamoyltransferase Activity in Human Serum
Open Access
- 1 May 1975
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 21 (6) , 754-757
- https://doi.org/10.1093/clinchem/21.6.754
Abstract
Ornithine carbamoyltransferase (EC 2.1.3.3) activity is a sensitive, specific indicator of hepatocellular injury. This paper describes development of an improved automated procedure for measurement of this activity. Triethanolamine—ethylenedlaminetetraacetate is used as a buffer, and activity is determined by measuring the concentration of the product, citrulline. Kinetic studies have been performed to determine optimal pH and L-ornithine and carbamoyl phosphate concentrations. Recovery of citrulline was studied. The upper limit of normal obtained in a study of 106 blood-bank donors was 6 U/liter. The automated procedure developed as a result of these studies, in which optimal assay conditions are used, produces a threefold increase in sensitivity and permits use of a sample volume of 1 ml.Keywords
This publication has 0 references indexed in Scilit: