Factors influencing the estimation of extravascular lung water by transpulmonary thermodilution in critically ill patients*
- 1 June 2005
- journal article
- research article
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 33 (6) , 1243-1247
- https://doi.org/10.1097/01.ccm.0000164566.23147.35
Abstract
To investigate factors that may influence the estimation of extravascular lung water (EVLW) with a single (cold) indicator compared with assessment using two indicators (thermo-dye dilution). Post hoc analysis of an electronic hemodynamic database. Surgical intensive care unit of a university hospital. Forty-eight critically ill patients monitored by the thermo-dye dilution technique in the postoperative period. None. The EVLW was simultaneously assessed by the thermo-dye dilution technique (EVLWref) and estimated by transpulmonary thermodilution (EVLWest). EVLWref index ranged between 1 and 40 mL/kg (mean 10 +/- 7 mL/kg) and EVLWest between 2 and 39 mL/kg (mean 9 +/- 6 mL/kg). EVLWref was closely correlated (r = .96) with EVLWest. The mean difference (bias) between EVLWref and EVLWest was -0.5 +/- 1.9 mL/kg. The bias was not influenced by the weight, height, body surface area, body mass index, Pao2, intrathoracic blood volume, cardiac output, or dosage of vasoactive agents. In contrast, the bias was slightly but significantly influenced by EVLWref, Pao2/Fio2 ratio, tidal volume, and level of positive end-expiratory pressure. In our surgical intensive care unit population, the estimation of EVLW by transpulmonary thermodilution was influenced by the amount of EVLW, the Pao2/Fio2 ratio, the tidal volume, and the level of positive end-expiratory pressure. However, compared with the double indicator method, transpulmonary thermodilution estimation remained clinically acceptable even in patients with severe lung disease.Keywords
This publication has 22 references indexed in Scilit:
- Accuracy of transpulmonary thermodilution versus gravimetric measurement of extravascular lung waterCritical Care Medicine, 2004
- Global End-Diastolic Volume as an Indicator of Cardiac Preload in Patients With Septic Shock *Chest, 2003
- Right ventricular function and positive pressure ventilation in clinical practice: from hemodynamic subsets to respirator settingsIntensive Care Medicine, 2003
- Assessment of cardiac preload and extravascular lung water by single transpulmonary thermodilutionIntensive Care Medicine, 2000
- Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patientsIntensive Care Medicine, 1999
- Effects of catecholamines on the pulmonary venous bed in sheepCritical Care Medicine, 1998
- Reduction in pulmonary blood volume during positive end-expiratory pressureJournal of Surgical Research, 1983
- Effect of lung inflation and hypoxia on pulmonary arterial blood volumeJournal of Applied Physiology, 1977
- On the Theory of the Indicator-Dilution Method for Measurement of Blood Flow and VolumeJournal of Applied Physiology, 1954
- The Dye Dilution Method for Describing the Central CirculationCirculation, 1951