Three‐dimensional T1ρ‐weighted MRI at 1.5 Tesla
Open Access
- 19 May 2003
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 17 (6) , 730-736
- https://doi.org/10.1002/jmri.10296
Abstract
Purpose: To design and implement a magnetic resonance imaging (MRI) pulse sequence capable of performing three‐dimensional T1ρ‐weighted MRI on a 1.5‐T clinical scanner, and determine the optimal sequence parameters, both theoretically and experimentally, so that the energy deposition by the radiofrequency pulses in the sequence, measured as the specific absorption rate (SAR), does not exceed safety guidelines for imaging human subjects.Materials and Methods: A three‐pulse cluster was pre‐encoded to a three‐dimensional gradient‐echo imaging sequence to create a three‐dimensional, T1ρ‐weighted MRI pulse sequence. Imaging experiments were performed on a GE clinical scanner with a custom‐built knee‐coil. We validated the performance of this sequence by imaging articular cartilage of a bovine patella and comparing T1ρ values measured by this sequence to those obtained with a previously tested two‐dimensional imaging sequence. Using a previously developed model for SAR calculation, the imaging parameters were adjusted such that the energy deposition by the radiofrequency pulses in the sequence did not exceed safety guidelines for imaging human subjects. The actual temperature increase due to the sequence was measured in a phantom by a MRI‐based temperature mapping technique. Following these experiments, the performance of this sequence was demonstrated in vivo by obtaining T1ρ‐weighted images of the knee joint of a healthy individual.Results: Calculated T1ρ of articular cartilage in the specimen was similar for both and three‐dimensional and two‐dimensional methods (84 ± 2 msec and 80 ± 3 msec, respectively). The temperature increase in the phantom resulting from the sequence was 0.015°C, which is well below the established safety guidelines. Images of the human knee joint in vivo demonstrate a clear delineation of cartilage from surrounding tissues.Conclusion: We developed and implemented a three‐dimensional T1ρ‐weighted pulse sequence on a 1.5‐T clinical scanner. J. Magn. Reson. Imaging 2003;17:730–736.Keywords
This publication has 32 references indexed in Scilit:
- Proteoglycan‐induced changes in T1ρ‐relaxation of articular cartilage at 4TMagnetic Resonance in Medicine, 2001
- The role of relaxation times in monitoring proteoglycan depletion in articular cartilageJournal of Magnetic Resonance Imaging, 1999
- Proton T1ρ‐dispersion imaging of rodent brain at 1.9 TJournal of Magnetic Resonance Imaging, 1998
- T1ρ MR imaging characteristics of human anterior tibial and gastrocnemius musclesAcademic Radiology, 1998
- T1ρ dispersion imaging of head and neck tumors: A comparison to spin lock and magnetization transfer techniquesJournal of Magnetic Resonance Imaging, 1997
- Spin lock and magnetization transfer imaging of head and neck tumors.Radiology, 1996
- Myocardial suppression in vivo by spin locking with composite pulsesMagnetic Resonance in Medicine, 1996
- T1ρ dispersion imaging of diseased muscle tissueThe British Journal of Radiology, 1993
- Spin locking for magnetic resonance imaging with application to human breastMagnetic Resonance in Medicine, 1989
- A Method for Tlp ImagingJournal of Computer Assisted Tomography, 1985