Reduction of susceptibility artifact in gradient‐echo imaging
- 1 January 1992
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 23 (1) , 193-200
- https://doi.org/10.1002/mrm.1910230120
Abstract
A new technique with which susceptibility artifact in gradient‐echo imaging can be reduced substantially by use of a tailored RF pulse is described. The proposed technique can ideally be applied to the case where high local magnetic field inhomogeneity is dominated by the susceptibility. The signal loss and void phenomena due to susceptibility in a voxel are studied and a correction method is also proposed. The description of the tailored RF pulse and its proposed application are given and experimental results obtained using a human volunteer with a 2.0‐T KAIS NMR system are presented. © 1992 Academic Press, Inc.Keywords
This publication has 15 references indexed in Scilit:
- Missing pulse steady-state free precessionMagnetic Resonance in Medicine, 1989
- Variable-rate selective excitationJournal of Magnetic Resonance (1969), 1988
- Direct FLASH MR imaging of magnetic field inhomogeneities by gradient compensationMagnetic Resonance in Medicine, 1988
- An analysis of fast imaging sequences with steady‐state transverse magnetization refocusingMagnetic Resonance in Medicine, 1988
- Chemical‐shift imaging with large magnetic field inhomogeneityMagnetic Resonance in Medicine, 1987
- Transverse coherence in rapid FLASH NMR imagingJournal of Magnetic Resonance (1969), 1987
- Rapid Fourier imaging using steady‐state free precessionMagnetic Resonance in Medicine, 1987
- Clinical Magnetic Susceptibility Mapping of the BrainJournal of Computer Assisted Tomography, 1987
- Use of frequency‐modulated radiofrequency pulses in MR imaging experimentsMagnetic Resonance in Medicine, 1986