Microprocessor detection of electrocardiogram R-waves
- 1 January 1979
- journal article
- Published by Taylor & Francis in Journal of Medical Engineering & Technology
- Vol. 3 (5) , 235-241
- https://doi.org/10.3109/03091907909160662
Abstract
ECG monitoring devices for coronary care units require reliable R-wave detectors as a prerequisite for subsequent analysis. A simple algorithm has been developed for the detection of R-waves in an ECG signal with a microprocessor (Motorola MC6800). Detection criteria used were gradient and duration of a single upslope or downslope. Rejection of muscle noise and other artifacts was achieved by setting system constants to optimum values which were determined by a short preliminary trial. This constant threshold system has been shown to cope with sudden changes of QRS complex amplitude better than some self adjusting systems. Assessment runs using patient tapes from a coronary care unit show total errors due to false positives and negatives are typically about 1% of the actual number of R-wave occurrences.Keywords
This publication has 5 references indexed in Scilit:
- Evaluation of computer diagnosis of ectopic beats encountered in routine patient monitoringComputers in Biology and Medicine, 1975
- COMPARISON OF ARRHYTHMIA COMPUTER AND CONVENTIONAL MONITORING IN CORONARY-CARE UNITThe Lancet, 1975
- Online analysis of e.c.g. arrhythmias with a digital computerMedical & Biological Engineering & Computing, 1975
- A low-cost general-purpose analogue QRS wave detectorMedical & Biological Engineering & Computing, 1974
- A QRS Preprocessor Based on Digital DifferentiationIEEE Transactions on Biomedical Engineering, 1971